Error | Cause | Remedy |
Unstable temperature baseline measurement | - Rapid cooling or warming of patient
| - Wait for stabilisation phase
|
| | - Stop volume supply if possible
|
| - Arterial connection cable broken or not correctly connected
| - Check cable connection replace as necessary
|
| - Faulty thermistor on PiCCO catheter
| - Check temperature change when flushing catheter, replace catheter if no temperature change seen
|
| | - Change position, change possibly patient positioning
|
Thermodilution curve is not displayed | - Incorrect injection at central venous catheter
| - Perform central venous injection correctly
|
| - Faulty thermistor on PiCCO catheter
| - Check temperature change when flushing catheter, replace catheter if no temperature change seen
|
| - Broken or failed cable connection of arterial connection cable
| - Check cable connection, change if necessary
|
Thermodilution curve display is too late (Time out limit)
| | |
| - Very low cardiac output or very high lung water
| - Use colder and/or more injectate volume
|
Thermodilution curve very flat
| - Inappropriate scale settings
| |
| | |
| - Very low cardiac output or very high lung water
| - Use colder and/or more injectate volume
|
Thermodilution curve displays more than one peak
| | |
| - Existing Intracardiac Shunts
| |