Tips & Tricks for haemodynamic monitoring

Arterial pressure measurement

Error

Cause

Remedy

 

Pressure curve absent

  • Pressure line not correctly attached

  • Check pressure line, stop-cock, cable and plug

  • Pressure line blocked

  • Flush pressure line

  • Cable or device defect

  • Connect pressure transducer directly to the bedside monitor for testing

Abnormal pressure curve

  • Catheter malpositioned

  • Check or change catheter position

  • Blocked catheter lumen

  • Flush catheter

  • Arrhythmia, extra systole

  • Regulate
    therapeutically

  • Tachycardia
    (> 240 bpm)

  • Regulate
    therapeutically

Flat pressure curve

  • Inappropriate scale settings

  • Set scale of
    pressure curve

  • Failed connection of pressure system

  • Check and ensure pressure system connected correctly

  • Flush has insufficient pressure

  • Increase the flush pressure to 250-300 mmHg

  • None or wrong zero adjustment

  • Carry out / repeat zero adjustment

Damped pressure transducer transmission from:

  • Airbubbles

  • Modifications of pressure line
  • Malposition of stop cocks
  • Kinked catheter or pressure line

 

 

  • Disconnect from patient and remove air bubble
  • Remove any additional lines and or stop cocks
  • Adjust stop cocks correctly
  • Remove kinks, possibly change patient position

  • Hypotonia

  • Regulate therapeutically

Indikator injection

Error

Cause

Remedy

 

Injection was not detected

  • CVC malpositioned

  • Correct position

  • 3-way-stopcock incorrectly adjusted

  • Adjust 3-way-stopcock

  • Temperature sensor housing pin clotted

  • Change temperature sensor housing

  • Temperature sensor housing / sensor not connected correctly

  • Check connection („click-sound“)

  • Defective temperature sensor

  • Check or change sensor

Thermodilution results

Error

Cause

Remedy

 

Results implausible



  • Amount injected does
    not match amount to be
    injected displayed by device

  • Inject correct volume

Pulse contour results

Error

Cause

Remedy

Calibration
PCCO not possible

  • Arterial pressure curve absent or faulty

  • Proof / optimise
    arterial pressure measurement

Large variation in values

  • Arrhythmia, extra systoles

  • Regulate therapeutically

Thermodilution curve

Error

Cause

Remedy

Unstable temperature baseline measurement

  • Rapid cooling or warming of patient

  • Wait for stabilisation phase

  • Massive volume loading

  • Stop volume supply if possible

  • Arterial connection cable broken or not correctly connected

  • Check cable connection replace as necessary

 

  • Faulty thermistor on PiCCO catheter

  • Check temperature change when flushing catheter, replace catheter if no temperature change seen

 

  • Parietal thermistor

  • Change position, change possibly patient positioning

Thermodilution curve is not displayed

  • Incorrect injection at central venous catheter

  • Perform central venous injection correctly

 

  • Faulty thermistor on PiCCO catheter

  • Check temperature change when flushing catheter, replace catheter if no temperature change seen

 

  • Broken or failed cable connection of arterial connection cable

  • Check cable connection, change if necessary


Thermodilution curve display is too late (Time out limit)

  • Injection too slow

  • Inject faster
    (< 7s)

  • Very low cardiac output or very high lung water

  • Use colder and/or more injectate volume


Thermodilution curve very flat

  • Inappropriate scale settings

  • Adjust scaling

 

  • Injectate too warm

  • Use colder injectate

 

  • Very low cardiac output or very high lung water

  • Use colder and/or more injectate volume


Thermodilution curve displays more than one peak

  • Injection not smooth

  • Ensure smooth injection

  • Existing Intracardiac Shunts

  • Shunt diagnosis

 